Moldex Healthcare Particulate Respirators/ Surgical Mask User Instructions


A. APPLICATIONS
  1. WARNING TO USER 
    1. Non-latex straps are used in the headbands of these respirators. However, individuals highly sensitive to natural rubber latex may have an allergic reaction.
    2. This product does not eliminate the risk of contracting any disease or infection. Change immediately if the respirator comes in contact with blood or fluids, including body fluids.
    3. OSHA has not set a permissible exposure level for airborne biohazards.
    4. This product has not been sold with warning or use instructions for personnel involved in industrial or related situations. If you are considering such uses, first contact the Moldex Technical Services Department at +1 (800) 421-0668 or +1 (310) 837-6500, ext. 512/550.

  2. RESTRICTIONS

    1. Before use, a written respiratory protection program in accordance with 29 CFR 1910.134 must be implemented.
    2. This respirator does not supply oxygen and must not be used in atmospheres containing less than 19.5% oxygen.
    3. Prior to each use, carefully inspect the entire respirator, including filter media and strap attachment area for tears and damage. Staple or strap attachment perforations do not effect NIOSH approval.
    4. If respirator is damaged, distorted, a proper fit cannot be obtained or breathing becomes difficult, leave contaminated area as soon as possible and replace respirator.
    5. If the respirator comes in contact with blood or fluids, including body fluids, leave contaminated area as soon as possible and discard and replace the respirator.
    6. Do not alter, modify or abuse this respirator.
    7. Store unused respirators in box/bag in a clean, dry, non-contaminated area.
    8. Dispose of respirator according to your employer’s policy and local regulations.
    9. Do not reuse or store for reuse or hang around neck unless your employer specifically authorizes reuse.
    10. When used for surgical procedures, discard after every use.
    11. Use respirator before the “use by” expiration date printed on box/bag.

      For technical assistance call Moldex Technical Service Department, +1 (800) 421-0668 or +1 (310) 837-6500, ext. 512/550.

  3. INTENDED USE

    • The various models of Moldex Type N95 Healthcare Particulate Respirators and Surgical Masks meet CDC Guidelines for TB Exposure Control within healthcare facilities. These devices are also intended to be worn by healthcare personnel during surgical procedures to protect both the patient and healthcare personnel from the transfer of microorganisms, body fluids and particulate material.

  4. CONTRAINDICATIONS

    1. Not for use with beards, or other facial hair that prevents direct contact between the face and sealing surface of the respirator.
    2. Eyewear must not prevent direct contact between the face and sealing surface of the respirator.
    3. Not to be used on children.

  5. DESCRIPTION
  • The Moldex Healthcare Series N95 Particulate Respirator and Surgical Mask is designed to help provide respiratory protection for the wearer. This product has been tested1 and certified by NIOSH as an N95 respirator and as having a filter efficiency level of 95% or greater against particulate aerosols free of oil. It is fluid resistant,2 disposable and may be worn in surgery or throughout the hospital.
  1. Tested in accordance with NIOSH 42 CFR 84.
  2. Passed ASTM F 1862 @ 160 mm Hg
B. APPROVALS
  1. Review NIOSH/DHHS label in approvals section of disposable respirator box or bag.